OBJECTIVE. To assess the performance and limitations of a reprocessing protocol for nonlumen electrophysiology catheters by testing the sterility of reprocessed devices and defining the maximum number of reprocessing cycles sustainable by the device in hygienically safe conditions. DESIGN. Simulated use, reprocessing, and testing of the catheters. SETTING. Microbiology and virology department of a public health diagnostic laboratory. INTERVENTIONS. Seventy-three catheters were collected after clinical use on patients. The first group of devices was tested for sterility after 1 cycle of reprocessing. By the repetition of simulated use (blood inoculated with bacteria) and reprocessing (decontamination, cleaning, and hydrogen peroxide gas plasma sterilization), we obtained 39 sample devices reprocessed 2 times, 26 reprocessed 3 times, 28 reprocessed 4 times, 36 reprocessed 5 times, and 22 reprocessed 6 times. Devices were cultured for 28 days in trypticase soy broth. RESULTS. We tested 208 catheters with 6 cycles of reprocessing and 4 inoculated bacteria species. No devices tested positive for the inoculated strains until the fourth cycle of reprocessing. One of 35 catheters showed the growth of the inoculated strain Bacillus subtilis after 5 cycles of reprocessing, and 1 of 22 catheters showed growth of this organism 6 cycles. After the second reprocessing, 7 of 36 devices showed growth of gram-negative bacteria other than the strain inoculated. CONCLUSIONS. Reprocessing according to the reprocessing protocol was insufficient to guarantee device sterility after 5 reuses. Cleaning with enzymatic solution revealed good cleaning properties with efficient bioburden reduction. Storage intervals of longer than 24 hours during reprocessing should be avoided to limit contamination or bacterial overgrowth. Technical considerations suggest the introduction of reprocessing procedures only in hospitals with considerable workloads. © 2006 by The Society for Healthcare Epidemiology of America. All rights reserved.

Sterility and microbiological assessment of reused single-use cardiac electrophysiology catheters

Tessarolo, Francesco;Nollo, Giandomenico
2006-01-01

Abstract

OBJECTIVE. To assess the performance and limitations of a reprocessing protocol for nonlumen electrophysiology catheters by testing the sterility of reprocessed devices and defining the maximum number of reprocessing cycles sustainable by the device in hygienically safe conditions. DESIGN. Simulated use, reprocessing, and testing of the catheters. SETTING. Microbiology and virology department of a public health diagnostic laboratory. INTERVENTIONS. Seventy-three catheters were collected after clinical use on patients. The first group of devices was tested for sterility after 1 cycle of reprocessing. By the repetition of simulated use (blood inoculated with bacteria) and reprocessing (decontamination, cleaning, and hydrogen peroxide gas plasma sterilization), we obtained 39 sample devices reprocessed 2 times, 26 reprocessed 3 times, 28 reprocessed 4 times, 36 reprocessed 5 times, and 22 reprocessed 6 times. Devices were cultured for 28 days in trypticase soy broth. RESULTS. We tested 208 catheters with 6 cycles of reprocessing and 4 inoculated bacteria species. No devices tested positive for the inoculated strains until the fourth cycle of reprocessing. One of 35 catheters showed the growth of the inoculated strain Bacillus subtilis after 5 cycles of reprocessing, and 1 of 22 catheters showed growth of this organism 6 cycles. After the second reprocessing, 7 of 36 devices showed growth of gram-negative bacteria other than the strain inoculated. CONCLUSIONS. Reprocessing according to the reprocessing protocol was insufficient to guarantee device sterility after 5 reuses. Cleaning with enzymatic solution revealed good cleaning properties with efficient bioburden reduction. Storage intervals of longer than 24 hours during reprocessing should be avoided to limit contamination or bacterial overgrowth. Technical considerations suggest the introduction of reprocessing procedures only in hospitals with considerable workloads. © 2006 by The Society for Healthcare Epidemiology of America. All rights reserved.
File in questo prodotto:
File Dimensione Formato  
2006 Tessarolo et al Sterility and Microbiological Assessment of Reused.pdf

non disponibili

Descrizione: Full paper
Tipologia: Documento in Post-print
Licenza: NON PUBBLICO - Accesso privato/ristretto
Dimensione 159.88 kB
Formato Adobe PDF
159.88 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11582/308049
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
social impact