One of the critical issues while operating bolometric detectors over periods of time of 1 year or more consists of keeping their response stable within a 0.1% level, despite the unavoidable temperature fluctuations of the cryogenic set-up. By using an energy pulser, which periodically delivers a fixed amount of energy in the absorber, it is possible to stabilize the response of the bolometers. A stabilization technique using heating devices, made up of heavily doped semiconductor material (well above the metal-to-insulator transition), has been developed in the framework of the CUORE experiment. In this paper we describe in detail the procedure for the realization of the heating elements, based on silicon planar technology. We then report on the multi-step low temperature characterization (77 K, 4.2 K, 1.5 K, 35 mK) of the heaters. Finally, an example of achieved stabilization for a CUORE-like detector is reported. The ∼1500 heaters tested at View the MathML source show less than 0.5% change in resistance between View the MathML source and 3 mV, and less than 1% change in value between 50 mK and 800 mK. In particular, the resistance change between 4.2 K and 1.5 K is less than 0.1%.
Production, characterization, and selection of the heating elements for the response stabilization of the CUORE bolometers
Margesin, Benno;
2012-01-01
Abstract
One of the critical issues while operating bolometric detectors over periods of time of 1 year or more consists of keeping their response stable within a 0.1% level, despite the unavoidable temperature fluctuations of the cryogenic set-up. By using an energy pulser, which periodically delivers a fixed amount of energy in the absorber, it is possible to stabilize the response of the bolometers. A stabilization technique using heating devices, made up of heavily doped semiconductor material (well above the metal-to-insulator transition), has been developed in the framework of the CUORE experiment. In this paper we describe in detail the procedure for the realization of the heating elements, based on silicon planar technology. We then report on the multi-step low temperature characterization (77 K, 4.2 K, 1.5 K, 35 mK) of the heaters. Finally, an example of achieved stabilization for a CUORE-like detector is reported. The ∼1500 heaters tested at View the MathML source show less than 0.5% change in resistance between View the MathML source and 3 mV, and less than 1% change in value between 50 mK and 800 mK. In particular, the resistance change between 4.2 K and 1.5 K is less than 0.1%.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.